15

7. Prepare a sufficient volume of detection antibodies (1x) using

detection antibody diluent. Each well requires 2.5 µl of detection
antibodies (10x) adjusted to a final volume of 25 µl with detection
antibody diluent (refer to the example below).

8. After incubating the samples, slowly remove and discard the sealing

tape, then

vacuum-filter.

9.

Wash 3 times.

10. Vortex the detection antibodies gently and add 25 µl to each well.

Incubate for 30 min.

11. While the detection antibodies are incubating, perform a 30 sec

quick-spin centrifugation of the streptavidin-PE (100x) prior to
pipetting to collect the entire volume at the bottom of the vial.

12. Prepare a sufficient volume of streptavidin-PE (1x) using assay buffer.

Each well requires 0.5 µl of streptavidin-PE (100x) adjusted to a final
volume of 50 µl with assay buffer (refer to the example on the
following page).

Example Detection Antibody Calculations

# of Wells

10x Detection

Antibody (µl)

Total Volume (µl)

96

48

32

24

300

150

100

75

2,700

1,350

900

675

3,000

1,500

1,000

750

Detection Antibody

Diluent (µl)