Bio-Rad Bio-Plex Pro Human Angiogenesis Reagent and Diluent Kits User Manual

Page 29

Advertising
background image

27

Possible Causes

Possible Solutions

High Background Signal

Incorrect buffer was used
(for example, assay buffer
used to dilute standards)

Use sample matrix or serum
standard diluent to dilute
the standards.

Spiked “0 pg/ml” wells by mistake

Be careful when spiking standards.
Do not add any antigens in the
0 pg/ml (blank) point.

Streptavidin-PE incubated
too long

Follow the procedure incubation
time.

Poor Recovery

Expired Bio-Plex reagents were
used

Check that reagents have not
expired. Use new or unexpired
components.

Incorrect amounts of components
were added

Check your calculations and be
careful to add the correct volumes.

Microplate shaker set to an
incorrect speed

Check the microplate shaker speed
and use the recommended setting.
Setting the speed too high may
cause splashing and contamination.
Use the recommended plate shaker.

Pipetting technique

Pipet carefully when adding
standards, samples, detection
antibodies, and streptavidin-PE,
especially when using a
multichannel pipet. Use a calibrated
pipet. Change pipet tip after every
volume transfer.

Advertising
This manual is related to the following products: