1. Select Calibrate

and confirm that the default values for

CAL1 and CAL2 are the same as the values on the Bio-Plex
calibration bead labels. Use the Bio-Plex low RP1
target value.

2. Select OK and follow the instructions for CAL1 and CAL2

calibration.

Bio-Plex Manager Software version 4.0, 4.1 and 4.1.1
This software versions permit calibration at either low or high
RP1 target values. The selection of which target value to use is
assay dependent. For Bio-Plex Pro angiogenesis assays,
calibrate using the low RP1 target value on the Cal 2 bottle. Re-
calibration is necessary between assays requiring different RP1
target values. Daily calibration is recommended before acquiring
data.

1.

Select Calibrate

and confirm that the default values

for CAL1 and CAL2 are the same as the values on the Bio-
Plex calibration bead labels. Use the Bio-Plex low RP1
target value.

2.

Select OK and follow the instructions for CAL1 and CAL2

calibration.

Prepare Protocol
Lot-specific assay protocols are available for download at www.bio-
rad.com/bio-plex/standards. If the desired protocol is not available for
download, create a new protocol.

1. Open a new protocol by selecting File, then New from the main

menu. Locate the steps at the left of the protocol menu.

2. Select Step 1 (Describe Protocol) and enter information about the

assay.

3. Select Step 2 (Select Analytes) and choose the appropriate

angiogenesis panel name, if available in the menu. If the panel name
is not available, create a new panel: