Possible Causes

Possible Solutions

Low Bead Count

Miscalculation of bead dilution

Check your calculations and be
careful to add the correct volumes.

Beads clumped in multiplex
bead stock tube

Vortex for 15–20 sec at medium
speed before aliquoting beads.

Vacuum on for too long when
aspirating buffer from wells

Do not apply vacuum to the filter
plate for longer than 10 sec after the
buffer is completely drained from
each well.

Did not shake filter plate enough
before incubation steps and prior
to reading

Shake the filter plate at 1,100 rpm
for 30 sec before incubation steps
and immediately before reading
the plate.

Reader is clogged

Refer to the troubleshooting guide
in the Bio-Plex hardware
instruction manual.

Low Signal or Poor Sensitivity

Standards reconstituted incorrectly Follow the standard preparation

instructions carefully (section 7).

Detection antibody or
streptavidin-PE diluted incorrectly

Check your calculations and be
careful to add the correct volumes.