Description:
Sinergy 1KVA UPS, 120 V AC input/output, 60 Hz line 
frequency. 5-minute backup time. 
Ordering:
Uninterruptible Power Supply, 120 V
(catalog # A30-10010)
One kVA UPS provides clean and uninterruptible power 
for the US and countries utilizing 120 V AC, 60 Hz power. 
This unit will provide 5–8 minutes of backup time in the 
event of a utility power loss.
We recommend the following UPS for countries utilizing 230 V AC, 
50 Hz power:
Model:
S1000XA-2 from ONEAC (available from Bio-Rad)
Description:
Sinergy 21KVA UPS, 230 V AC input/output, 50 Hz 
line frequency. 5-minute backup time.
Ordering:
Uninterruptible Power Supply, 230 V
(catalog # A30-10020)
One kVA UPS provides clean and uninterruptible 
power for countries utilizing 230 V AC, 50 Hz power. 
This unit will provide 5–8 minutes of backup time in 
the event of a power loss.
Setting Up the 
ProteinChip
SELDI System
Setting Up the ProteinChip SELDI System
If you are planning to put the reader onto your network please refer to 
Appendix A
Required materials
•
ProteinChip SELDI reader
•
Workstation computer and associated peripherals
•
Crossover cable (included)
Connect the reader and the workstation
1.
After the ProteinChip SELDI system reader has been powered up, allow up 
to 2 hours for the vacuum pump to reach optimum pressure.
2.
Plug the crossover cable into the network slots on the reader, and into the 
computer. The crossover cable is clearly labeled (on the cable itself).
3.
Power up the computer.