Warranty and Service Agreements

Page 23

14. Replace the thumb screws on the front of the pawl drive.

15. Manually open and close the load lock door to verify proper operation and

sealing. Leave the load lock door in the closed position.

16. Replace the rear cassette guide.

17. Reconnect the “stack full” sensor cable, then replace the reader front cover.

18. Turn on the reader power.

19. Verify that the reader initializes correctly.

20. Insert a cassette and insert an array to verify proper load lock door operation.

Load Lock Door Cleaning for Personal Edition Readers

1.

Open the sample exchange door on the front of the reader.

2.

Using the pins on the front, unscrew the load lock door from its post
(counterclockwise).

3.

Clean the load lock O-ring and the flat side of the door with a clean lab
wipe and alcohol.

4.

Replace the load lock door by screwing it clockwise back on to the post.

5.

Tighten the door until it seals against the O-ring.