Mass resolving power, Mass accuracy, Laser – Bio-Rad ProteinChip Data Manager Software User Manual

Page 7: Sample handling, Vacuum system

Advertising
background image

Page 2

ProteinChip SELDI System: Reader Guide

Mass Resolving Power

Peptide

Better than 1000 for [Arg

8

]-vasopressin at 1084 Da, standard operating voltage

Mass Accuracy

Inherent

Measured mass will be within 0.5% of true mass with 95% confidence
(0.25% CV)

External calibration

Measured mass will be within 0.1% of true mass with 95% confidence
(0.05% CV)

Internal calibration

Measured mass will be within 0.01% of true mass with 95% confidence
(0.005% CV)

Laser

Pulsed nitrogen laser (337 nm), 4 ns pulse width

Sample Handling

Personal

Runs one 8-spot array within 20 minutes using 250 transients per spot

Enterprise

Runs a cassette of up to twelve 8-spot ProteinChip arrays within 2 hours using
250 transients per spot

Vacuum System

Operating pressure < 2.7 x 10

-4

Pa

Gauges: active inverted magnetron (high vacuum) and active Pirani (low
vacuum) turbomolecular pump