Multicassette autoloader (enterprise edition), Overview of operation – Bio-Rad ProteinChip Data Manager Software User Manual

Page 19

Advertising
background image

Page 14

ProteinChip SELDI System: Reader Guide

Multicassette Autoloader (Enterprise Edition)

The ProteinChip SELDI system Enterprise Edition comes with a multicassette
autoloader. It contains cassette loader and unloader assemblies designed to
allow the automated processing of multiple cassettes from a stack. The purpose
of this feature is to allow unattended operation of the reader for significant
periods of time. The loader processes the cassettes sequentially, removing each
cassette from an input stack, cycling the designated arrays in the cassette
through the load lock and source chamber, then transferring the processed
cassette to the output unloader before returning to collect the next cassette. This
system is designed for continuous unattended processing of up to 14 cassettes
in one run. A maximum of 14 cassettes can be loaded into the stacker at one
time. The cassette lids must be closed in order for the reader to run.

Overview of Operation

The following instructions are intended to be a brief guide only. For more
information on data acquisition and analysis, please consult the ProteinChip data
manager software manual.

Acquiring control of the reader

The status of the sensors is displayed in the Home tab of the software interface
for the reader (refer to Fig. 12). The sensors should have the following status
before beginning operation:

Enterprise Edition

Vacuum state — operating

High voltage — off

Cassette — present

Loader cover — closed

Unloader cover — closed

High vacuum — <2.7 x 10

-6

Pa.

Control of the reader is obtained by clicking the Start icon.

Personal Edition

High voltage — off

Door — closed

Lock — locked

Control of the reader is obtained by clicking the Start icon.

Advertising