Dako Pascal User Manual

Page 13

Advertising
background image

Pascal Handbook

13

Section 4

Operating Instructions

Set Up

1. Familiarize yourself with Pascal and its parts prior to operating.

2. Adjust the pan by matching the pan handles with the body handles.

Note If the outside of the pan is wet, Pascal will make a cracking noise. Any water in the
body may cause malfunctions.

3. Fill the pan with 500 mL of deionized water for performing the standard Target Retrieval
procedure.

4. Ensure the heat shield is in the center of the pan.

5. Place 1 to 4 Tissue-Tek

®

containers each filled with 250 mL of suitable Target Retrieval

Solution in the pan. Alternatively, use plastic Coplin jars. Fill each plastic Coplin jar with
50 mL of Target Retrieval Solution. Place slides intended for the Target Retrieval
procedure in the Tissue-Tek

®

containers or Coplin jars.


6. Place a dry Pascal Quality Strip on top of the containers.

7. Position the lid on the pan and secure in place by aligning the OPEN arrow etched on
the lid’s surface with one of the white dots on the pan handle. Grip the lid handle and
rotate clockwise so that the CLOSE arrow is aligned with the white dot on the pan handle.

8. Place the weight (petcock) on the vent nozzle located on the Pascal chamber lid.

9. Turn the red main power switch to the ON position (I).

Programming

Pascal Factory Settings

Push the Display Set button to recall the following factory setting parameters on a new
instrument. These settings can be changed for individual applications. Refer to Changing
Factory Setting within this section for notes on making changes for specific applications.

Pascal Factory Settings

Factory Settings

Temperature

Time

SP1 125

ºC 30

seconds

SP2

90 ºC

10 Seconds

SP Limit

10 ºC

n/a