Pascal Calibration Protocol


Pascal calibration is performed at two temperatures. Both calibration points must be
performed for a complete Pascal calibration procedure. While the complete calibration
procedure is recommended every 6 months, it is also recommended to perform the room
temperature calibration prior to each use

Room Temperature Calibration

1.

Plug Pascal into power supply
• Turn on the power switch located on the right side, next to the front panel.


2.

Place 500 mL water into Pascal pan and equilibrate to room temperature for 15

20 minutes.
• This step is necessary in order to validate the starting temperature.

• Do not add warm water to speed up the process.


3.

Read the temperature on the digital display.


4.

Insert a calibrated thermometer in the Pascal pan containing 500 mL of water.
Compare the thermometer temperature reading to the digital display reading. If the
digital display matches the thermometer reading within 0.5 ºC then the instrument
should be considered within calibration limits for the Room Temperature
Calibration check point. Proceed to the High Temperature Calibration.

If the readings fail to match by greater than 0.5 ºC, adjust the digital reading by
following these steps:
• Turn the main power switch off.
• Turn it back on while pressing the Up

and Down

arrows at the same

time.

• Confirm that all 6 LEDs are illuminated; this indicates that Pascal is in

calibration offset mode.

• Press the “Display Set” button one time.
• Use the Up

or Down

arrow to adjust the offset so the digital

temperature read out will match the calibrated thermometer. Example: The
digital reading is 25 ºC, and the thermometer reading is 26 ºC. The action
required is to adjust the offset up by 1 ºC.

• Once the calibration is set press the “Display set” button two times to exit the

calibration mode.


High Temperature Calibration

1.

Ensure that Pascal is set to factory settings for SP1, SP2, and SP Limit for the
High Temperature Calibration procedure. If Pascal settings have been changed
from factory specified settings refer to the Operating Instructions section in this