Dako Pascal User Manual

Page 7

Advertising
background image

Pascal Handbook

7

Connect Power

Pascal is a 110V–120V instrument. In locations with 110– 120 V supply power, simply plug
the instrument into the appropriate outlet.

If the location supplies 220–240 V power, a step down isolation transformer is required
(dual winding, rated 1000 VA minimum, 50–60 Hz, 220–240 V input 110–120 V output
nominal 13A, with an IEC female output connector). An isolation transformer can be
connected with portable cord and plug to the main supply. Pascal is then connected to the
isolation transformer.

Do not operate Pascal with other types of high electrical demand equipment such as
incubators, microwave ovens, or refrigerators on the same circuit. This may affect the
performance of the instrument.

The main power switch is red and is located to the right of the operation panel on the front
of the instrument. This switch will illuminate in the ON position.

I = ON and O = OFF