Mechanical Considerations

SureTemp Plus Module OEM Implementation

14

Confidential

Welch Allyn OEM Technologies

Figure 1. Probe Cover Box Receptacle

Mount the SureTemp Plus module within 30˚ of probe-well vertical to help prevent the probe from
falling out of the well during transport or movement of the finished device.

Allow for ventilation of the module and probe well to avoid excessive heat build-up. Add
ventilation slots for adequate air movement to keep the probe tip at ambient room temperature.

Consider ergonomics when designing the finished product to avoid damaging the probe tip
during removal and replacement into the well. The majority of probe problems stem from
mishandling the probe or cord. It is recommended that the probe connector be mounted so that
the probe cord is not stressed and does not interfere with other device accessories.

The host must integrate a mechanical lock for the probe connector. (Refer to the latch detail in
the illustration on page 18.)

note

Keep the probe and probe covers at ambient room temperature. Measurement accuracy
is negatively affected if the probe tip and the probe covers are not kept at ambient room
temperature.