2a35) set thermometer probe mode, Command code (hex), Data – Welch Allyn SureTemp PlusModule - User Manual User Manual

Page 43: Response data, Notes

Advertising
background image

SureTemp Plus Module OEM Implementation

Communications Protocol

Welch Allyn OEM Technologies

Confidential 43

(2A35) Set Thermometer Probe Mode

Causes the module to set the current probe mode from the host.

Command Code (hex)

2A35

Data

Size (bytes)

Type

Description

1

int8

0: Oral
1: Pediatric Axillary
2: Adult Axillary

Response Data

Size (bytes)

Type

Description

0

N/A

None

Notes

The probe mode controls whether the oral, pediatric axillary, or adult axillary algorithm is to
be used with an oral/axillary probe. This command has no effect if the current probe type is
rectal.