9416) send 690 event log, Command code (hex), Data – Welch Allyn SureTemp PlusModule - User Manual User Manual
Page 49: Response data, 941a) send programmable ambient bias

SureTemp Plus Module OEM Implementation
Communications Protocol
Welch Allyn OEM Technologies
Confidential 49
(9416) Send 690 Event Log
Causes the module to send between 0 and 10 bytes of data back to the host depending on
the number of errors in the event log. 10 is the maximum size of the event log, but if it is
partially empty then the response could be less then 10 bytes in length.
Command Code (hex)
9416
Data
Size (bytes)
Type
Description
0
N/A
None
Response Data
Size (bytes)
Type
Description
≤
10
uint8
All errors are 1-byte. See page 71 for a mapping
between log numbers and display number.
(941A) Send Programmable Ambient Bias
Causes the module to send the programmable ambient bias to the host.
Command Code (hex)
941A
Data
Size (bytes)
Type
Description
0
N/A
None
Response Data
Size (bytes)
Type
Description
4
float32
The device’s ambient bias. It is different in the OEM
applications and must be updated.