SureTemp Plus Module OEM Implementation

42

Confidential

Welch Allyn OEM Technologies

(2A12) Set Temperature Method

Causes the module to set the current temperature method from the host, forcing the module
into the host-specified mode.

Command Code (hex)

2A12

Data

Size (bytes)

Type

Description

1

int8

0: Normal (Predictive)
1: Monitor

Response Data

Size (bytes)

Type

Description

0

N/A

None

Notes

This command returns an “Invalid operation in this context” error if the temperature probe is
out of the probe well, and returns a “Data Range” error if the specified temperature method is
undefined.
This command has effect only in a predict cycle, and can be used only to transition the
monitor from predict mode to monitor mode.