8 - communications protocol, Introduction, Communications interface – Welch Allyn SureTemp PlusModule - User Manual User Manual

Page 30: System eeprom, Communications interface system eeprom

Advertising
background image

Introduction

SureTemp Plus Module OEM Implementation

30

Confidential

Welch Allyn OEM Technologies

8 - Communications Protocol

Introduction

The SureTemp Plus module is a command-driven slave device communicating with a host
system over an asynchronous serial communication line. The host initiates communications by
sending a command block; the SureTemp Plus module always responds with a response block.

The SureTemp Plus module and the host system are referred to below as “module” and “host”
respectively.

The protocol description includes three main sections:

Communications Interface

9600 baud rate
8 data bits, no parity, 1 stop bit
full duplex
Asynchronous using standard non-return to zero (NRZ) format
Non-isolated module power and communications are provided at J1

System EEPROM

The system EEPROM contains system configuration and operational constants that are encoded
by Welch Allyn. While the majority of parameters pertain to Welch Allyn manufacturing and
operational defaults, the OEM can customize some parameters, such as the following, to
accommodate requirements:

Packet Structure

describes the protocol

Commands and Responses

describes the commands

Command and Response Examples

provides examples

Thermometer Algorithm

The algorithm—oral or axillary—used when the oral probe is
installed. (Default: Oral)

Temperature Method

Predictive or monitor mode. (Default: Predictive)