(2A37) Set Default Thermometer Probe Mode
Causes the module to set the default probe mode from the host. This flag is used to define
what algorithm the device defaults to when an oral/axillary probe is connected to the device.
Command Code (hex)
2A37
Data
Size (bytes)
Type
Description
1
int8
0: Oral/Axillary
1: Pediatric Axillary
2: Adult Axillary
3: Use Last Algorithm
Response Data
Size (bytes)
Type
Description
0
N/A
None
Notes
Use the “Save Calibration Data to EEPROM” command defined in for (2A06) Save Data to
EEPROM on page 53 to save permanently to EEPROM.
(3000) Restart Module
Causes the module to reboot.
Command Code (hex)
3000
Data
Size (bytes)
Type
Description
0
N/A
None
Response Data
Size (bytes)
Type
Description
0
N/A
None