Introduction

SureTemp Plus Module OEM Implementation

26

Confidential

Welch Allyn OEM Technologies

5 - Verification Testing

Introduction

This section provides calibration verification procedures.

Calibration Verification

Testing Module Only

This procedure provides a convenient means of testing the thermometer module only. It requires
the use of the Welch Allyn SureTemp Plus Calibration Key.

Procedure

Follow the instructions on the Cal Key tag.

1. Extract the probe and disconnect it from the probe input connector.

2. Insert the SureTemp Plus Calibration Key into the probe connector receptacle and observe

the readings.

3. Insert a probe shaft into the probe well and then remove it to initiate a temperature-taking

cycle.

For a correctly calibrated module, the Cal Key value should read 97.3˚F ± 0.2˚F.

4. Remove the Calibration Key and reinstall the probe connector plug.

5. Return the probe into the well.

Testing the Module and the Probe

The Welch Allyn 9600 Calibration Tester (see page 29 for the appropriate part number) provides
a convenient means of testing the entire thermometer system, module and probe.

To test the system, follow the instructions provided with the 9600 Calibration Tester.

note

This Cal Key Test does not test the probe. To test the probe, use the 9600 Calibration
Tester.