5 - verification testing, Introduction, Calibration verification – Welch Allyn SureTemp PlusModule - User Manual User Manual
Page 26: Testing module only, Procedure, Testing the module and the probe, Introduction calibration verification, Testing module only procedure

Introduction
SureTemp Plus Module OEM Implementation
26
Confidential
Welch Allyn OEM Technologies
5 - Verification Testing
Introduction
This section provides calibration verification procedures.
Calibration Verification
Testing Module Only
This procedure provides a convenient means of testing the thermometer module only. It requires
the use of the Welch Allyn SureTemp Plus Calibration Key.
Procedure
Follow the instructions on the Cal Key tag.
1. Extract the probe and disconnect it from the probe input connector.
2. Insert the SureTemp Plus Calibration Key into the probe connector receptacle and observe
the readings.
3. Insert a probe shaft into the probe well and then remove it to initiate a temperature-taking
cycle.
For a correctly calibrated module, the Cal Key value should read 97.3˚F ± 0.2˚F.
4. Remove the Calibration Key and reinstall the probe connector plug.
5. Return the probe into the well.
Testing the Module and the Probe
The Welch Allyn 9600 Calibration Tester (see page 29 for the appropriate part number) provides
a convenient means of testing the entire thermometer system, module and probe.
To test the system, follow the instructions provided with the 9600 Calibration Tester.
note
This Cal Key Test does not test the probe. To test the probe, use the 9600 Calibration
Tester.