SureTemp Plus PC Display

SureTemp Plus Module OEM Implementation

78

Confidential

Welch Allyn OEM Technologies

SureTemp Plus PC Display

Figure 12. Handheld Window

The Handheld window closely emulates a Welch Allyn
SureTemp Plus Thermistor Thermometer.

The Handheld view is intended to be a suggestion or
recommendation as to how one should present the
temperature measurement, status, and advisory
messages when implementing the OEM module into a
new system.

Operation

For information on operating the Handheld, refer to
“SureTemp PLUS Handheld” in the SureTemp Plus PC
Help files.

1. Pull the probe from the probe well and verify

that all segments are shown on the simulated
LCD display.

2. Verify that the oral mode is selected (the

flashing head icon on the handheld display). If
this icon is not present, click the Mode
Selection button until the flashing head icon
appears.

3. Load a probe cover by inserting the probe into

a probe cover and pressing the probe handle
down firmly. The probe handle moves slightly to
engage the probe cover.

4. With the Oral Mode indicator displayed, quickly

place the probe tip under the patient’s tongue on either side of the mouth to reach the
rear sublingual pocket (Figure 13). Have the patient close his/her lips around the probe.

Figure 13. Location of the Rear Sublingual Pockets

rear sublingual pockets
(under the tongue)