Figure 3. state diagram: predict – Welch Allyn SureTemp PlusModule - User Manual User Manual

Page 17

Advertising
background image

SureTemp Plus Module OEM Implementation

OEM Implementation

Welch Allyn OEM Technologies

Confidential 17

Figure 3. State Diagram: Predict

Probe pulled;

Start Predict

Algorithm ran 1

or many cycles;

Active = 1

Algorithm determined that

probe tip is at desired temperature;

Signal Ready beep;

Initialize the in-progress indicator

No Predict

Idle

Monitor

No

Predict

Predict

Display

Final X

Not Ready

Ready

Predict

Predict

Position

Predict complete;

Display result

Algorithm time-out

Predict in progress;

Increment and display

the in-progress indicator

Predict started;

Reset the in-progress

indicator

Algorithm reset;

Start over

Tissue contact regained;

Reset the in-progress indicator

Tissue contact lost;

Display the out-of-position indicator

Algorithm not active;

Terminate the temperature cycle

Switch to Monitor mode

Algorithm reset;

Start over

(Algorithm Status = Final) & Active;

Display the temperature

FINAL

FINAL_OVER_TEMP
FINAL_OVER_TEMP_POSITION

FINAL_UNDER_TEMP
FINAL_UNDER_TEMP_POSITION

FINAL_POSITION

Display Predict Complete Temperature.

Do not show the result; indicate over-temperature.

Do not show the result; indicate under-temperature.

Show the final temperature with an out-of-position indication.

Final states:

Algorithm time-out

Advertising