Adult axillary mode, Pediatric axillary mode, Rectal mode – Welch Allyn SureTemp PlusModule - User Manual User Manual
Page 21

SureTemp Plus Module OEM Implementation
OEM Implementation
Welch Allyn OEM Technologies
Confidential 21
Adult Axillary Mode
Figure 6. Acceptable Ways to Display the Adult Axillary Mode
Pediatric Axillary Mode
Figure 7. Acceptable Ways to Display the Pediatric Axillary Mode
Rectal Mode
Preferred
To indicate the adult axillary mode, display an upright body icon with a
flashing upper body (Figure 6) and the words ADULT AXILLARY.
Acceptable
If a graphical display cannot be accomplished, display the words
ADULT AXILLARY or the abbreviation A AXILLARY, A ALY, A AXY,
A AX, A-AXILLARY, A-ALY, A-AXY, or A-AX.
Not acceptable
To provide no indication, or to use any other abbreviation or
inappropriate indicator, is not acceptable.
Preferred
To indicate the pediatric axillary mode, display a kneeling body icon
with a flashing arrow pointing to the axilla (Figure 7) and the words
PEDIATRIC AXILLARY.
Acceptable
If a graphical display cannot be accomplished, display the words
PEDIATRIC AXILLARY or the abbreviation P AXILLARY, P ALY,
P AXY, P AX, P-AXILLARY, P-ALY, P-AXY, or P-AX.
Not acceptable
To provide no indication, or to use any other abbreviation or
inappropriate indicator, is not acceptable.
Preferred
To indicate the oral mode, display an upright body icon with a flashing
lower body (Figure 8) and the word RECTAL.
Acceptable
If a graphical display cannot be accomplished, display the word
RECTAL or the abbreviation REC.
Not acceptable
To provide no indication, or to use any other abbreviation or
inappropriate indicator, or to use the word ANAL, is not acceptable.
A AXILLARY
A-AXILLARY
A ALY
A-ALY
A AXY
A-AXY
A AX
A-AX
P AXILLARY
P-AXILLARY
P ALY
P-ALY
P AXY
P-AXY
P AX
P-AX