Normal Mode
1. The program defaults to Normal (predict) mode upon initial operation; however, the mode
settings are remembered. With the probe in the well, select the Mode item from the menu
bar. Verify that none of the three mode choices is selected from the drop-down menu list.
The unit is now ready to display temperatures in Normal mode when the probe is
removed from the probe well.
2. The thermometer remains in Normal mode until another mode is selected (checked).
Monitor Mode
1. With the probe removed from the well, use the mouse to click on the Mode button on the
Handheld.
2. The thermometer remains in Monitor mode until the Monitor menu item is disabled
(unchecked).
note
Use only Welch Allyn probe covers. The use of other manufacturer’s probe covers or no
probe cover may produce temperature measurement errors and/or inaccuracy.
note
When using the Monitor mode, allow the temperature readout to stabilize for 3 minutes
for oral and rectal temperatures and 5 minutes for axillary temperatures.