9433) set mode pause for clinical device, Command code (hex), Data – Welch Allyn SureTemp PlusModule - User Manual User Manual
Page 51: Response data, Notes

SureTemp Plus Module OEM Implementation
Communications Protocol
Welch Allyn OEM Technologies
Confidential 51
(9433) Set Mode Pause For Clinical Device
Enables the host to set the number of milliseconds the device pauses after a predict
temperature before switching to monitor mode. The transfer to monitor mode occurs after the
host-specified delay.
This is appropriate during clinical testing when comparing a predicted temperature to a
monitored temperature. It makes the final predict measurement available for a period before
the module switches to monitored mode and begins sending new temperature values.
Command Code (hex)
9433
Data
Size (bytes)
Type
Description
2
int16
0 = Disabled
Do not automatically switch to monitor mode after a
predict; without a pause the user would never see
the final predicted temperature.
Response Data
Size (bytes)
Type
Description
0
N/A
None
Notes
Use the “Save Calibration Data to EEPROM” command (page 41) to save permanently to
EEPROM.