2945) send thermometer probe mode, Command code (hex), Data – Welch Allyn SureTemp PlusModule - User Manual User Manual
Page 40: Response data, 2949) send default thermometer probe mode

Commands and Responses
SureTemp Plus Module OEM Implementation
40
Confidential
Welch Allyn OEM Technologies
(2945) Send Thermometer Probe Mode
Causes the module to send the current thermometer probe mode to the host. This setting
defines the algorithm to be used the next time a temperature is taken with an oral/axillary
probe connected.
Command Code (hex)
2945
Data
Size (bytes)
Type
Description
0
N/A
None
Response Data
Size (bytes)
Type
Description
1
int8
0: Oral/Axillary
1: Pediatric Axillary
2: Adult Axillary
(2949) Send Default Thermometer Probe Mode
Causes the module to send the default thermometer probe mode setting. The default
thermometer probe mode defines the algorithm to be used when an oral/axillary probe is
connected to this device.
Command Code (hex)
2949
Data
Size (bytes)
Type
Description
0
N/A
None
Response Data
Size (bytes)
Type
Description
1
int8
0: Oral
1: Pediatric Axillary
2: Adult Axillary
3. Use last