2945) send thermometer probe mode, Command code (hex), Data – Welch Allyn SureTemp PlusModule - User Manual User Manual

Page 40: Response data, 2949) send default thermometer probe mode

Advertising
background image

Commands and Responses

SureTemp Plus Module OEM Implementation

40

Confidential

Welch Allyn OEM Technologies

(2945) Send Thermometer Probe Mode

Causes the module to send the current thermometer probe mode to the host. This setting
defines the algorithm to be used the next time a temperature is taken with an oral/axillary
probe connected.

Command Code (hex)

2945

Data

Size (bytes)

Type

Description

0

N/A

None

Response Data

Size (bytes)

Type

Description

1

int8

0: Oral/Axillary
1: Pediatric Axillary
2: Adult Axillary

(2949) Send Default Thermometer Probe Mode

Causes the module to send the default thermometer probe mode setting. The default
thermometer probe mode defines the algorithm to be used when an oral/axillary probe is
connected to this device.

Command Code (hex)

2949

Data

Size (bytes)

Type

Description

0

N/A

None

Response Data

Size (bytes)

Type

Description

1

int8

0: Oral
1: Pediatric Axillary
2: Adult Axillary
3. Use last