9413) send mode pause for clinical device, Command code (hex), Data – Welch Allyn SureTemp PlusModule - User Manual User Manual
Page 48: Response data, 9415) send default thermometer mode

Commands and Responses
SureTemp Plus Module OEM Implementation
48
Confidential
Welch Allyn OEM Technologies
(9413) Send Mode Pause For Clinical Device
Causes the module to send the number of milliseconds between the final predict temperature
and its transfer into monitor mode to the host. The host can control the transfer time from
predict mode to monitor mode.
Command Code (hex)
9413
Data
Size (bytes)
Type
Description
0
N/A
None
Response Data
Size (bytes)
Type
Description
2
int16
0 = Disabled; do not automatically switch to monitor
mode after a predict. This is because without a
pause the user would never see the final predicted
temperature.
Limits 0-10,000 where 10,000 = 10 seconds max.
(Value in milliseconds).
(9415) Send Default Thermometer Mode
Causes the module to send to the host the default thermometer mode. This allows the user to
specify which method of taking temperature readings we enter when the probe is pulled,
normal predictive mode, or monitor mode.
Command Code (hex)
9415
Data
Size (bytes)
Type
Description
0
N/A
None
Response Data
Size (bytes)
Type
Description
1
int8
0: Normal (Predictive)
1: Monitor