Welch Allyn SureTemp PlusModule - User Manual User Manual
Page 35

SureTemp Plus Module OEM Implementation
Communications Protocol
Welch Allyn OEM Technologies
Confidential 35
TEMP (33 data bytes)
Type
Description
Values/Units
float32
“Final/Predicted” temperature
in
°
Kelvin
float32
Current Probe Temperature
in
°
Kelvin
uint16
Current Heater Count/PWM
0-999
uint16
Sample Counter - 173 mS/sample
int8
Method
0: Normal
1: Monitor
2: Clinical
int8
Algorithm
0: Oral
1: Rectal
2: Pediatric Axillary
3: Cal Key
4: Adult Axillary
int8
SureTemp state
0: Idle
1: Not Ready
2: Ready
3: Predict
4: Predict display
5: Predict position
6: Final
7: Final position
8: Final over temp
9: Final over temp position
10: Final under temp
11: Final under temp position
12: No predict (monitor mode)
20: Error status
int16
Error code
0: No error
See Error Codes (page 65)
boolean
active
Probe out and Active
int8
Physical probe type connected
0: Oral/Axillary
1: Cal Key
2: Rectal
3: No Probe
float32
Current Probe Resistance
in Ohms
float32
ambient_temp
Ambient temp used in final result (Kelvin)
float32
ptb_resistance
PTB resistance read at beginning of a cycle
uint16
battery
Battery voltage in hundreds of volts.
e.g. 4.5 = 450. This is the battery voltage
actually given to the SureTemp algorithm.