SureTemp Plus Module OEM Implementation

Communications Protocol

Welch Allyn OEM Technologies

Confidential 35

TEMP (33 data bytes)

Type

Description

Values/Units

float32

“Final/Predicted” temperature

in

°

Kelvin

float32

Current Probe Temperature

in

°

Kelvin

uint16

Current Heater Count/PWM

0-999

uint16

Sample Counter - 173 mS/sample

int8

Method

0: Normal
1: Monitor
2: Clinical

int8

Algorithm

0: Oral
1: Rectal
2: Pediatric Axillary
3: Cal Key
4: Adult Axillary

int8

SureTemp state

0: Idle
1: Not Ready
2: Ready
3: Predict
4: Predict display
5: Predict position
6: Final
7: Final position
8: Final over temp
9: Final over temp position
10: Final under temp
11: Final under temp position
12: No predict (monitor mode)
20: Error status

int16

Error code

0: No error
See Error Codes (page 65)

boolean

active

Probe out and Active

int8

Physical probe type connected

0: Oral/Axillary
1: Cal Key
2: Rectal
3: No Probe

float32

Current Probe Resistance

in Ohms

float32

ambient_temp

Ambient temp used in final result (Kelvin)

float32

ptb_resistance

PTB resistance read at beginning of a cycle

uint16

battery

Battery voltage in hundreds of volts.

e.g. 4.5 = 450. This is the battery voltage
actually given to the SureTemp algorithm.