Saving Ambient Bias

SureTemp Plus Module OEM Implementation

72

Confidential

Welch Allyn OEM Technologies

45

Probe,
Module,
Patient

Measurement below allowable
temperature values and far below the
low ambient or patient limits.

Patient or environmental temperature
conditions may be too low. Verify
conditions at 80

°

F and 50

°

F

respectively. If conditions are valid and
problem persists, replace probe. If
problem still persists replace module.

46

Probe,
Module,
Patient

Measurement above allowable
temperature values and way beyond the
high ambient and/or patient limits.

Patient or environmental temperature
conditions may be too high. Verify
conditions at 110

°

F and 104

°

F

respectively. If conditions are valid and
problem persists, replace probe. If
problem still persists replace module.

47

Module

Internal calibration resistor (RCAL) on
the board is damaged.

If problem persists, replace module.

48

Module

Internal calibration resistor (RCAL) on
the board is damaged.

If problem persists, replace module.

49

Module

Internal circuit validation resistor (PTB)
on the board is damaged.

If problem persists, replace module.

50

Module

Internal circuit validation resistor (PTB)
on the board is damaged.

If problem persists, replace module.

51

Module

A/D measurement timed out and did not
finish in the allotted time slot.

If problem persists, replace module.

52

Probe

Probe not characterized/calibrated.

Replace probe.

59

Battery/
Voltage

Battery or power-supply voltage below
algorithm maximum value.

Verify that the supply voltage is within
specification.

60

Battery/
Voltage

Battery or power supply voltage
exceeds algorithm minimum value

Verify that the supply voltage is within
specification.

61

Module

Reference voltage circuit detected to be
under voltage or unstable.

If problem persists, replace module.

63

Module

Probe well missing or not installed
properly

Re-insert probe well or check for
alignment problem.

65

Module

Problem saving to the module's
EEPROM.

Verify that the supply voltage is
correct. Try to make changes again. If
problem persists, replace module.

66

Module

The module's or the probe's error
detection mechanism detected an error.

Verify that the supply voltage is
correct. Try again. If problem persists
replace indicated component.

67

Module

Problem reading the module's EEPROM
or saving to the module's EEPROM.

Verify that the supply voltage is
correct. Try to make changes again. If
problem persists, replace module.

68

Module

Problem reading the module's EEPROM
or saving to the module's EEPROM.

Verify that the supply voltage is
correct. Try to make changes again. If
problem persists, replace module.

69

Module

Problem reading the module's EEPROM
or saving to the module's EEPROM.

Verify that the supply voltage is
correct. Try to make changes again. If
problem persists, replace module.

Error Class

Possible Cause

Action