Welch Allyn SureTemp PlusModule - User Manual User Manual
Page 72

Saving Ambient Bias
SureTemp Plus Module OEM Implementation
72
Confidential
Welch Allyn OEM Technologies
45
Probe,
Module,
Patient
Measurement below allowable
temperature values and far below the
low ambient or patient limits.
Patient or environmental temperature
conditions may be too low. Verify
conditions at 80
°
F and 50
°
F
respectively. If conditions are valid and
problem persists, replace probe. If
problem still persists replace module.
46
Probe,
Module,
Patient
Measurement above allowable
temperature values and way beyond the
high ambient and/or patient limits.
Patient or environmental temperature
conditions may be too high. Verify
conditions at 110
°
F and 104
°
F
respectively. If conditions are valid and
problem persists, replace probe. If
problem still persists replace module.
47
Module
Internal calibration resistor (RCAL) on
the board is damaged.
If problem persists, replace module.
48
Module
Internal calibration resistor (RCAL) on
the board is damaged.
If problem persists, replace module.
49
Module
Internal circuit validation resistor (PTB)
on the board is damaged.
If problem persists, replace module.
50
Module
Internal circuit validation resistor (PTB)
on the board is damaged.
If problem persists, replace module.
51
Module
A/D measurement timed out and did not
finish in the allotted time slot.
If problem persists, replace module.
52
Probe
Probe not characterized/calibrated.
Replace probe.
59
Battery/
Voltage
Battery or power-supply voltage below
algorithm maximum value.
Verify that the supply voltage is within
specification.
60
Battery/
Voltage
Battery or power supply voltage
exceeds algorithm minimum value
Verify that the supply voltage is within
specification.
61
Module
Reference voltage circuit detected to be
under voltage or unstable.
If problem persists, replace module.
63
Module
Probe well missing or not installed
properly
Re-insert probe well or check for
alignment problem.
65
Module
Problem saving to the module's
EEPROM.
Verify that the supply voltage is
correct. Try to make changes again. If
problem persists, replace module.
66
Module
The module's or the probe's error
detection mechanism detected an error.
Verify that the supply voltage is
correct. Try again. If problem persists
replace indicated component.
67
Module
Problem reading the module's EEPROM
or saving to the module's EEPROM.
Verify that the supply voltage is
correct. Try to make changes again. If
problem persists, replace module.
68
Module
Problem reading the module's EEPROM
or saving to the module's EEPROM.
Verify that the supply voltage is
correct. Try to make changes again. If
problem persists, replace module.
69
Module
Problem reading the module's EEPROM
or saving to the module's EEPROM.
Verify that the supply voltage is
correct. Try to make changes again. If
problem persists, replace module.
Error Class
Possible Cause
Action