Typical Start-up Sequence

SureTemp Plus Module OEM Implementation

68

Confidential

Welch Allyn OEM Technologies

7. The module sends unsolicited <941C> Temperature Record packets as soon as the probe is

removed from the well.

8. The host monitors the state byte in the Temp packets. When the module goes from Not

Ready to Ready, the host can sound an audible tone or beep to signify that the oral/axillary
probe is ready to be inserted into the patient’s mouth. At this point, the probe is inserted. If
the probe is not inserted, the module remains in predict mode for one minute, and then
moves to the Monitor mode. It remains in Monitor mode until the cycle is inactive.

9. The host monitors temp state and error message bytes within the <941C> Temperature

Record packets.

10. A Final Predict Temperature is included by the module in the temp packet. The <941C> Temp

Record packets continue until the Probe is returned to the well or times out.