Mini-Rotary Viscometer

CMRV-5000 Instruction & Operation Manual

Version 1.0b—August, 2011; CANNON

®

Instrument Company

2139 High Tech Road • State College, PA • 16803 • USA

The profile may not be completed. Verify that the Remaining Time
for the profile is 0:00.

The instrument may not be communicating with the computer. Log
into VISCPRO as a Manager and click Service/MRV/Monitor to
open the MRV monitor. If the data window is updated regularly with
new data, then communication is being maintained.