Setting the cmrv instrument address, Gas purge option installation – Cannon Instrument CMRV-5000 User Manual

Page 20

Advertising
background image

14

CANNON

®

Mini-Rotary Viscometer

CMRV-5000 Instruction & Operation Manual

Version 1.0b—August, 2011; CANNON

®

Instrument Company

2139 High Tech Road • State College, PA • 16803 • USA

Address selection knob

CAUTION

Do NOT plug or unplug the pulleywheel assembly when the instrument
power is ON! To disconnect the sensor, pull it out by the knurled portion of
the plug.

Firmware installation notes

Download button

Current firmware is installed on your instrument during factory check-
out. However, the CMRV-5000 is capable of receiving firmware updates
from the computer.

When installing firmware, place the instrument in the firmware down-
load mode by pressing the DOWNLOAD button adjacent to the serial
port on the rear of the Controller. Follow the software prompts of the
firmware installation program to upgrade your instrument.

NOTE

If the DOWNLOAD button is pressed inadvertently, switch off the CMRV-
5000 power for at least four seconds, then
restore power to the unit.

Setting the CMRV instrument address

When installing/connecting a new CMRV
instrument, you must set the instrument
address using the SELECT ADDRESS
dial on the rear of the CMRV Controller.
This dial offers 16 settings (0-9, A-F).

Procedure

To set the address, rotate the dial (see photo) to a setting not currently in
use by other CMRV instruments.

CAUTION

The MRVW software will not function correctly unless each networked
CMRV instrument is set to a different address (see APPENDIX D for
multi-unit configuration).

Gas Purge option installation

If you have purchased the Gas Purge option for your CMRV instrument,
locate the Gas Purge kit, including the gas purge regulator, connecting
hardware, installation mounting bracket, and quick-connect gas purge
line. Obtain a Phillips screwdriver and follow the installation instructions
below:

1. Remove the two "plug" screws on the center of the left side panel of

the head unit (as viewed from the front of the unit). The screws are
located between the grillwork and are aligned vertically approxi-
mately 82.5 mm (3.25") apart.

2. Align the mounting bracket of the gas purge assembly with the screw

holes so that the connecting hardware on the attached regulator faces
the rear of the head unit. Secure with the Phillips-head screws provided.