Service menu options

To access the service options, click Service from the VISCPRO

®

primary display and select the desired instrument.

MRV Monitor

The MRV Monitor window permits the user to view current operational
values for the CMRV instrument. These settings may be helpful to
CANNON

®

technical personnel when troubleshooting difficulties with

the CMRV.

Cleaning Service

Select Cleaning Service to open the Cleaning Service window. Then
Click the Start Cleaning button to initiate an instrument heating cycle
(50°C). The current instrument temperature and instrument status will be
displayed in the window. Click the Stop Cleaning Service button when
cleaning is completed.

Temperature Probe Calib.

The Temperature Probe Calibration window provides a software
interface for calibrating the CMRV block temperature probe. Consult
Chapter 7 of the CMRV Instruction & Operation Manual for details
regarding the software-driven probe calibration procedure.