Measuring apparent viscosity – Cannon Instrument CMRV-5000 User Manual

Page 73

Advertising
background image

67

CANNON

®

Mini-Rotary Viscometer

CMRV-5000 Instruction & Operation Manual

Version 1.0b—August, 2011; CANNON

®

Instrument Company

2139 High Tech Road • State College, PA • 16803 • USA

11. Repeat steps 1-10 to perform yield stress/viscosity tests for all

remaining cells to be tested in order from left to right.

Notes on yield stress testing

 Step 9 (see previous section) is required for ASTM D 3829.

 The ASTM D 4684 method is a pass-fail method for fresh oils. If the

rotor fails to move with the weight of the 10-gram hook-cage, the oil
has failed the test. However, the MRVW program permits adding of
additional weights. If this alternate method is preferred, continue with
the procedure as described in step 9. Otherwise, click the Abort Test
button to abort yield stress testing on that cell.

NOTE

Do not abort CMRV temperature control until you have finished all yield
stress and viscosity tests for the cells.

 Pulley-wheel motion is detected by the digital-optical sensor. Ad-

equate rotor movement is defined as 13° of rotation for ASTM D
4684 and ASTM D 6821 (This is equivalent to 3 mm of motion by
the crossarm). For ASTM D 3829, 10° of rotation is required.

 If yield stress testing on a cell is aborted before adequate rotor

movement is detected, the resultant calculated yield stress will be
based on the highest weight used. For example, if the operator aborts
a yield stress calculation after testing with a total of 30 grams with
no rotor movement, the CMRV software will report yield stress
greater than 105 Pascals (Y>105).

Measuring apparent viscosity

After the yield stress test has concluded, you may measure apparent
viscosity using the procedure below:

1. Place the pulley-wheel in line with the appropriate viscometric cell

and place the rotor thread over the pulley-wheel.

NOTE

ASTM methodology requires testing the samples in sequence (l-r).

2. Ensure that a Sample ID has been entered for each sample. Then

click

for the desired sample. OR use the Tab key to highlight

the desired viscosity test icon and then press R. The Viscosity

LED (light-emitting diode) on the CMRV-5000 will glow and the

Start Test button will begin flashing rapidly.

3. Carefully place the hook-weight on the end of the thread (see photo,

next page), supporting the weight with your hand to prevent rotation
of the rotor.

5. Press the flashing Start Test button and proceed immediately to step 6.

6. Lift the locking pin and release the weight by lowering your hand

quickly and smoothly.

Advertising