weight are provided. Weights vary depending
on the ASTM Method used for testing.

Temperature probe

CMRV-5000 block temperature is detected by
a 3k-ohm temperature probe, which must be
seated securely in the thermistor aperture at
the side of the CMRV housing and plugged
into the electronic chassis.

Thermometers

The CMRV-5000 unit is shipped from CAN-
NON

®

with two thermometers to be used to

check the temperature of the block.

 The long thermometer used for the probe

and cell calibrations is a PRINCO mer-
cury-thallium calibrated thermometer with
a range of -46°C to +30°C in .2°C increments and an accuracy of ±
0.01°C below 20°C.

 The shorter, high-temperature uncalibrated thermometer has a range

of 0°C to 105°C in 1°C increments with an accuracy of ± 1°C.

NOTE

If a thermometer is removed from
the CMRV-5000 when the unit is
cooling, plug the thermometer
aperture in the Plexiglas

®

cover to

prevent ice formation around the
rotors.

Cell caps

Cell caps are used to enhance
temperature control and prevent
condensation in the CMRV cells
during the cooling phase of the
temperature profile.

Front panel operations

CMRV front panel

Control features

CMRV front panel control features are simple and functional. The switch
on the front panel is the power switch for the unit. The light-emitting

Hook-weights

Cell caps