Cannon Instrument CMRV-5000 User Manual

Page 45

Advertising
background image

39

CANNON

®

Mini-Rotary Viscometer

CMRV-5000 Instruction & Operation Manual

Version 1.0b—August, 2011; CANNON

®

Instrument Company

2139 High Tech Road • State College, PA • 16803 • USA

9. From the Calibrate window, click the desired cell number for calibra-

tion.

10. Press the blinking Start Test button on the CMRV Controller.

11. Raise the locking pin quickly and smoothly. The weight will apply

torque to the rotor. The computer will record the time for three
complete revolutions.

12. Permit the wheel to make at least three complete revolutions (the

viscosity LED will go out). Then catch and remove the weight before
the thread is completely unwound. Do not permit the thread to
completely unwind.

NOTE

After the rotor revolution time has been recorded, the program will
calculate the calibration constant for that cell.

13. At the conclusion of the test, the VISCPRO® software will display

the newly-calculated calibration constants.

NOTE

If the new constants vary from the previous cell constants by more than 4
percent, an error will be generated. It is generally advisable to recalibrate
the cell in this instance.

NOTE

To view previously-stored calibration constants, click Configure/MRV/
View Cell Constants.

14. Repeat the above calibration routine for the remaining cells.

Limiting calibration time

It is recommended that you perform the cell calibration immediately
following the one-hour sample soak time.

NOTE

If any cell has a calibration constant more than 10 percent higher or lower
than the average for the other cells, the fault may be excessive friction in
the upper bearing, improper seating of the lower bearing, or a damaged
rotor tip. Examine the rotor for damage and, if necessary, replace the
rotor and recalibrate the cell.

Saving calibration data

15. At the conclusion of the calibration, review the displayed data in the

Calibrate Cell Constants window. If the data is not acceptable,
recalibrate as necessary.

16. When acceptable values have been obtained, click Abort Tempera-

ture Control. Then click Save Data to save the new calibration
constants. Do not omit this step or new calibration data will be lost.

17. Close the Calibrate Cell Constants window and then click Main/

Save Instrument. The Save Instrument window will appear.

18. In the Save As box, input a name for the new instrument configura-

tion (which includes the newly-calculated calibration constants).
Then click OK to save the configuration.