APPENDIX D —

MULTI-UNIT CONFIGURATION

Introduction

The CMRV-5000 instrument have been designed to permit multi-unit
configurations of up to four CMRV units. Configuration is accomplished
via a “daisy-chain” technique using RS-485 connections.

A multi-unit interface kit containing all required
items for multi-unit configuration is available
from CANNON

®

(Catalogue # 9728-R40 for

120V units; #9728-R45 for
240V units).

Placement of multiple units

The practical “hands-on”
realities of viscosity and yield
stress testing suggest orienta-
tion of CMRV units in close
proximity with the computer
controlling the testing. Appro-
priate cables
(8' lengths) are provided with
the interface kit for this
purpose.

Procedure

Using the interface kit provided by CANNON

®

, complete the multi-unit

configuration as follows:

1. Make sure the power is switched off to the computer and all instru-

ments to be configured.

2. Locate the RS-232 to RS-485 converter provided with the interface

kit.

3. Plug the converter’s RS-232 25-pin connector into the COM port

you selected for multi-unit configuration during installation of the
CMRV software. If you are using COM 1, a DB-25 to 9-pin adaptor
may be required.

4. Select a CMRV unit as the first in the “daisy chain” and attach the

CAT-5 cable to the RS-485 connection marked IN on the first CMRV
chassis (see diagram, next page):