3-6

3

3 Window Functions and Operation

(3) Basic point

If several peaks exist when using the transparent object mode, this control specifies

whether to use the peak that is NEAR or FAR from the sensor head as the reference.

(4) ABLE

Sets whether to use AUTO or MANUAL control of the light intensity.

Normally set to "AUTO".

If set to "MANUAL", set the upper and lower limits for the light intensity control range

(setting range: 1 to 99).

(5) ABLE tuning

This function uses "ABLE" to tune the "MANUAL" setting range according to the actual

target. Refer to "Executing ABLE tuning" (page 3-7) for details.

(6) Reflection type selection mode

Sets the installation method to diffused reflection or specular reflection.

(7) Median

Sets the number of sampling points to use for the median filter.

(8) LASER control group

Sets the LASER control group.

(9) Range

Sets the range.

(10) Mask setting

Sets whether to enable (ON) or disable (OFF) masking.

(11) Mask boundary

If the "Mask setting" is set to "ON", this control specifies the mask boundary using POS-1 or

POS-2.

If POS-1 < POS-2: masks around POS-1 and POS-2.

If POS-1 > POS-2: masks between POS-1 and POS-2.

If POS-1 = POS-2: no masking is applied.

(12) Copy the head setting/Paste the copied head setting

Use this button to copy the head settings to another head (or a head in a different program

number). Refer to "Copying the head settings to another head" (page 3-7) for details.