KEYENCE TM-3000 Series User Manual

Page 76

Advertising
background image

A-8

App

end

ix

Error Messages

- Setup software TM-H1 TM-Navigator User’s Manual -

Program name

setting

The length of the program name exceeds the

setting range.

Enter a name that is no longer than 16

characters long.

Set the program name to 16 or fewer

characters.

2-3

Environment

settings

The IP address exceeds the setting range.

Set the value between 0 and 255.

Set the IP address to groups of numbers

between 0 and 255.

5-21

The subnet mask exceeds the setting range.

Set the value between 0 and 255.

Set the subnet mask to groups of numbers

between 0 and 255.

5-21

The gateway exceeds the setting range.

Set the value between 0 and 255.

Set the gateway to groups of numbers

between 0 and 255.

5-21

The strobe time exceeds the setting range.

Set the value between 2 and 100.

Set the strobe time between 2 and 100.

5-21

View

measurement

value

Failed to acquire measurement.

Set the controller to "Measurement screen

(during measurement)".

-

View image

Image capture failed.

Perform "Trigger input" with the controller to

capture the image.

4-6

No image is displayed.

Capture the image to TM-Navigator.

4-5

Data storage

The amount of storage data exceeds the setting

range.

Set the value between 1 and 65536.

Set the number of data points for data storage

between 1 and 65536.

4-14

Change unit settings to data storage.

Change unit settings to data storage.

-

There is no data to delete.

Store measurement data in the controller.

4-13

There is no data to read.

Store measurement data in the controller.

4-13

Image storage

Change unit settings to Image storage.

Change unit settings to Image storage.

-

There is no data to delete.

Store image data in the controller.

4-20

There is no data to read.

Store image data in the controller.

4-20

Communication

process

Failed to communicate.

• Set the correct communication parameters

for the computer and selected controller.

• Set the controller to "Measurement screen

(during measurement)".

1-7

-

Help

Cannot open HELP file.

Install Adobe Reader on the computer.

-

Item

Display contents

Remedy

Reference

page