External input, Deflt usr, External input setting screen – KEYENCE SK-1000 Series User Manual

Page 15: Zero shift input, Reset input, Timing input, Bank a input/bank b input, E sk-1000 im, Ztiming chart z minimum input time, Ztiming chart

Advertising
background image

15

E SK-1000 IM

10. External input

You can select functions to be assigned to four external inputs (External input 1 to

4).

„

External input setting screen

When the user setting is set, each function of the external input 1 to 4 can be set.
Functions that can be set differ for each external input. (When the default is set,

these four screens are skipped.)

Press S / T button on each setting screen to specify functions

Reference

When the same function is selected for multiple external inputs, the

inputs function according to OR operation.

„

Zero shift input

When the zero shift input is set to ON, the internal measurement value (R.V.) is

shifted (offset) to the selected shift target value.

z

Timing chart

z

Minimum Input Time

ON time: 20 ms

OFF time: 20 ms

„

Reset input

When the reset input is turned ON while using the hold function, the judgment

value (P.V.) in the held state is canceled and the value becomes [-----].

z

Timing chart

z

Minimum Input Time

ON time: 20 ms

OFF time: 20 ms

T1 = Input response delay (20 ms max.) + reset time*

* Reset time (judgment output, analog output) = Sampling period × (averaging

rate + 2)

External input

Description

Default

value

Default setting

Functions for the external inputs remain the same
as the factory default setting.

c

User setting

Functions for the external inputs are changed from
the factory default setting.

External Input 1 (Pink wire)

External Input 2 (Yellow wire)

Setting

value

Description

Default

value

Setting

value

Description

Default

value

Shift

Zero shift input

c

reset

Reset input

c

bnK-a

Bank A input

bnK-a

Bank A input

bnK-b

Bank B input

bnK-b

Bank B input

off

Not used

off

Not used

External input 3 (Pink/Purple wire)

External Input 4 (Purple wire)

Setting

value

Description

Default

value

Setting

value

Description

Default

value

tiMnG

Timing Input

c

bnK-a

Bank A input

c

bnK-a

Bank A input

bnK-b

Bank B input

bnK-b

Bank B input

off

Not used

off

Not used

10. in

dEFLt

Usr

HOLD

HI

V

kV

ALARM

BANK

ZERO SHIFT

TIMING

TEMP

%RH

HI

LO

R.V.

0

1

2

3

LO

GO

HOLD

HI

V

kV

ALARM

BANK

ZERO SHIFT

TIMING

TEMP

%RH

HI

LO

R.V.

0

1

2

3

LO

GO

HOLD

HI

V

kV

ALARM

BANK

ZERO SHIFT

TIMING

TEMP

%RH

HI

LO

R.V.

0

1

2

3

LO

GO

HOLD

HI

V

kV

ALARM

BANK

ZERO SHIFT

TIMING

TEMP

%RH

HI

LO

R.V.

0

1

2

3

LO

GO

HOLD

HI

V

kV

ALARM

BANK

ZERO SHIFT

TIMING

TEMP

%RH

HI

LO

R.V.

0

1

2

3

LO

GO

HOLD

HI

V

kV

ALARM

BANK

ZERO SHIFT

TIMING

TEMP

%RH

HI

LO

R.V.

0

1

2

3

LO

GO

10.External Input

External input 1

setting screen

External input 2

setting screen

External input 3

setting screen

External input 4

setting screen

11.Bank switching

method

[MODE]

or X

[MODE]

or X

[MODE]

or X

[MODE]

or X

[MODE]

or X

Internal measurement
value (R.V.)

Max. 20 ms

Min. 20 ms

Min. 20 ms

R.V. before shift

R.V. after shift

ON

OFF

Zero shift input

Judgment value
(P.V.) judgment
output

T1

Min. 20 ms

Min. 20 ms

Before resetting

[- - - - - ]

ON

OFF

Reset input

„

Timing Input

When the timing input is turned ON while using the hold function, the judgment
value (P.V.) is held.

z

Timing chart

When Sample hold is set
"Timing Input" (page 15) setting: "
Level"

"Timing Input" (page 15) setting: "Edge"

When Peak hold, Bottom hold or Absolute value hold is set
"Timing Input" (page 15)
setting: "Level"

"Timing Input" (page 15) setting: "Edge"

z

Minimum Input Time

ON time: 2 ms
OFF time: 2 ms

„

Bank A input/Bank B input

By combining the ON/OFF status of bank A input and bank B input, you can
switch between banks.

z

Timing chart

Reference

When the judgment value (P.V.) is held using the hold function, the
judgment value (P.V.) remains held even if the ON/OFF status of bank A
input or bank B input is switched. However, the judgment output will
change according to the HI/LO setting value of the new bank.

z

Minimum Input Time

ON time: 20 ms
OFF time: 20 ms

Measuring period

Measuring period

Max. 2 ms

Max. 2 ms

Max. 2 ms

Hold released

Sample hold

Min. 2 ms

Min. 2 ms

ON

OFF

Timing Input

Judgment value
(P.V.) judgment
output

Judgment value
(P.V.) judgment
output

Max. 2 ms

Max. 2 ms

Max. 2 ms

Sample hold

Sample hold

Sample hold

Min. 2 ms

Min. 2 ms

ON

OFF

Timing Input

Max. 2 ms

Max. 2 ms

Min. 2 ms

Min. 2 ms

ON

OFF

Max. 2 ms

Sampling does not start Sampling period

Sampling starts

Previous result

Judgment value
(P.V.) judgment
output

Timing Input

Judgment value
(P.V.) judgment
output

Max. 2 ms

Max. 2 ms

Min. 2 ms

Min. 2 ms

ON

OFF

Timing Input

Sampling period 1

Sampling period 1 result

Sampling period 2 result

Previous result

Sampling period 2

Sampling period 3

Judgment value
(P.V.) judgment
output

Max. 8 ms

Max. 20 ms

Max. 20 ms

Min. 20 ms

ON

OFF

Bank A input

ON

OFF

Bank B input

Max. 20 ms

Bank 1

Bank 2

Bank 3

Bank 0

In this example, after 8 ms have passed, Bank A
and Bank B are judged as OFF and operation at
Bank 0 may be performed.

Advertising