Safety precautions, General precautions, Safety information for the sk-1000 series – KEYENCE SK-1000 Series User Manual

Page 2: Abnormal conditions, Precautions for use, Other precautions, Precautions for regulations and standards

Advertising
background image

2

E SK-1000 IM

Safety Precautions

General Precautions

At startup and during operation, be sure to monitor the functions and
performance of this product and confirm normal operation.

We recommend that you take substantial safety measures to avoid any damage
in the event that a problem occurs.

If the product is modified or used in any way other than those described in the
specifications, its functions and performance cannot be guaranteed.

When this product is used in combination with other devices, the functions and
performance may be weakened, depending on the operating conditions,
surrounding environment, etc.

Do not use this product for the purpose of protecting the human body.

Do not subject this device, including peripheral devices, to rapid temperature
change. Product failure may occur due to condensation.

Safety Information for the SK-1000 series

Warming up

Leave the SK Series about 30 minutes after turning on the power.
The circuit is not stable immediately after the power turns on, so the display value
may gradually fluctuate.

Abnormal Conditions

Precautions for Use

Installation environment

To use this product normally and safely, do not install this product in the following

locations. Product failure may occur.

High-humidity, dusty or poorly-ventilated locations

High-temperature locations where the unit is exposed to direct sunlight

Locations where there is corrosive or combustible gas

Locations where the unit may be directly subjected to vibration or impact

Locations where water, oil or chemicals may splash onto the unit

Locations that cause the unit to discharge its static electricity

Influence of dirt

Measurement errors may occur due to dust, water, oil, etc.

Anti-noise prevention

When the unit is installed near electric noise source such as a power source or

high-voltage line, operational errors or product failure may occur. Take adequate

measures such as using a noise filter, separating cables or insulating the sensor

amplifier and the sensor head.

WARNING

This product is only intended to detect the charge on object(s).
Do not use this product for the purpose to protect a human body
or a part of human body.

This product is not intended for use as explosion-proof product.
Do not use this product in hazardous location and/or potentially
explosive atmosphere.

NOTICE

If the following conditions occur, turn OFF the power immediately.
Continuing to use this product under abnormal conditions may
cause product failure.

When water or foreign matter enters the SK Series

When the SK Series is dropped or the case is damaged

If smoke or unpleasant odor is present.

CAUTION

Use with the correct power source and voltage. Otherwise, fire,
electric shock or product failure may result.

Do not attempt to open or modify the SK Series. Doing so may
cause fire or electric shock.

NOTICE

Before disconnecting the cables, make sure to turn off the main
unit and devices connected to the main unit. Otherwise, the unit
may be damaged.

Do not turn off the power while modifying settings. Some or all
of the setting data may be lost.

The measurement area is basically a guide for measurement. A
proper size of target with respect to the measurement area is
recommended to ensure a more accurate measurement.

Power ON Reset

After the power is turned ON, it will take approx. 3.5 seconds for the measurement

to start. The judgment results will be output after the sampling rate has elapsed.

Other Precautions

Power source

Noise superimposed on the power supply may cause malfunction. Use a direct
current stabilized power source which uses an insulation transformer.

When using a commercially available switching regulator, make sure to ground
the frame ground terminal.

„

Grounding precautions

A grounding wire is present in the I/O cable of the SK-1000 Series.

Precautions for Regulations and

Standards

„

CE Marking

Keyence corporation has confirmed that this product complies with the essential
requirements of the applicable EC Directives, based on the following
specifications.
Be sure to consider the following specifications when using this product in the
Member States of European Union.

z

EMC Directive

Applicable standards

EN61326-1, ClassA

This product is intended to be used in an industrial electromagnetic
environment.

Remarks
These specifications do not give any guarantee that the end-product with this
product incorporated complies with the essential requirements of EMC Directive.
The manufacturer of the end-product is solely responsible for the compliance on
the end-product itself according to EMC Directive.

„

FCC regulations

This product complies with the following EMI regulations specified by the FCC.

Applicable standards

FCC part 15 subpart B, class A digital device

„

IC (Industry Canada) regulations

This product complies with the following EMI regulations specified by IC.

Applicable standards

ICES-003, class A digital apparatus

NOTICE

For proper measurements, be sure to connect the blue wire
protruding from the SK-1000 Series to ground.

Use class-D grounding-resistance not exceeding 100 Ω.

Advertising