Operation Theory

• 37

Timebase clock source

In scan acquisition mode, all the A/D conversions start on the output of
counters, which use Timebase as the clock source. With the software you can
specify the Timebase to be either an internal clock source (on board 24MHz) or
an external clock input on pin 45 of J1. The external clock is useful when you
want to acquire data at rates not available with the internal A/D sample clock.
The external clock source must generate TTL-compatible continuous clocks,
and the maximum frequency is 24MHz while the minimum is 1MHz.

Acquisition_in_progress

Scan_start

AD_conversion

Scan_in_progress
(SSH_OUT)(pin47)

3 Scans, 4 Samples per scan
(SC_Counter=3, DIV_Counter=4)

Sampling Interval t=
SI2_COUNTER/TimeBase

Scan Interval T=
SI_COUNTER/TimeBase

Ch2

Ch3

Ch1

Ch0

Ch2

Ch3

Ch1

Ch0

Ch2

Ch3

Ch1

Ch0

( channel sequences are specified in Channel Gain Queue)

Figure 9: Scan Timing

In the scan acquisition mode, the channel, gain, and input configurations
(single-end, differential, or U_CMMD) must be specified in the hardware
Channel Gain Queue, please refer to 5.1.3.2 for more details.

There are 5 trigger modes to start the scan acquisition, please refer to 5.1.3.3
details. The data transfer modes will be discussed in 5.1.3.4.