EXFO IQS-12002B Calibration System User Manual
Page 100
Configuring the FTB-7000D OTDR on the IQS-12002B
96
IQS-12002B
Creating the GP-120963: Singlemode Dead-Zone Calibration Instrument
1c. Enter the inventory number and comments if necessary.
Note: When the basic model of the calibration instrument is configured to appear
in the Calibration Certificate produced by the IQS-12002B, the information
that you enter in the Inventory number text box will appear in the section
entitled Standards used to establish traceability.
1d. Select Calibrated (no recalibration required).
2. Click OK.
3. From the Add a Calibration Event window, enter the date as written
on the instrument calibration certificate, and click OK.
4. Use the length of the reference standard given on the acceptance test
and add 3 meters to it (new length = 3 m + given length).