3 setting up your calibration system, Configuring areas specific for instruments, 3setting up your calibration system – EXFO IQS-12002B Calibration System User Manual
Page 23
19
3
Setting Up Your Calibration
System
Configuring Areas Specific for Instruments
Since the system can house instruments and DUTs in the same expansion
unit, you must specify which area is dedicated to the instruments. The
application will then be able to identify which modules must be
considered as instruments and which as DUTs.
To configure the area for instruments:
1. From the Tools menu, select System > Configure IQS Platform.
2. In the Instruments maximum address (CLUS) box, enter the address
of the last slot you want to use for instruments. The application will
assign the next slots to DUTs.
3. Click OK to confirm.
Do not modify the first
two numbers
Unit number
(indicated on front panel of
controller or expansion unit)
Slot number
Do not modify
Increase when DUT is
connected to the unit
by RS
-
232