EXFO IQS-12002B Calibration System User Manual
Page 38
Advertising
Operating Your Calibration System
34
Performing a Test
14. Back in the Measurement and Adjustment Procedure Selection box,
make sure Custom Specifications appears in the Specification
revision list.
15. Click Next and click Complete to close the wizard.
The application will prompt you to confirm if you want to perform the
acquisition now or later.
16. If the required standards have not been added to the database yet, the
application will prompt you to add them; proceed as follows:
16a.Click OK.
16b.Enter the required information and click OK.
16c. Confirm the authority name and calibration date.
16d.Click
and link the calibration certificate file provided by the
calibration authority.
16e.Click OK.