Configuring the FTB-7000D OTDR on the IQS-12002B

Calibration System

89

Creating the GP-121392: Distance Reference Standard

Creating the GP-121392: Distance Reference

Standard

This section presents the procedure to create a distance reference
standard.

To create the reference standard:
1.
From the Add a standard window, enter the relevant information:

1a. Match the calibration instrument appearing in the Calibration

Instrument list to the instrument you intend to create.

The calibration name will be displayed in the Calibration
Instrument
list.

1b. Enter the instrument serial number.

1c. Enter the inventory number and comments if necessary.

Note: When the basic model of the calibration instrument is configured to appear

in the Calibration Certificate produced by the IQS-12002B, the information
that you enter in the Inventory number text box will appear in the section
entitled Standards used to establish traceability.

1d. Select Calibrated (no recalibration required).