Cleaning the FOA Connector Adapter
Cleaning the FOA Connector Adapter
Make sure that the FOA connector adapter is properly cleaned to assure its
accuracy during calibration.
To clean your FOA connector adapter:
1. Remove the FOA from the instrument.
2. Slowly insert and gently rotate a cleaning tip that has been dipped in
isopropyl alcohol into the FOA.
3. Insert a dry cleaning tip to dry.
4. Blow away any remaining lint with clean compressed air.
5. Use a metal wire (diameter smaller than 150 μm) and carefully insert it
through the sleeve until it comes the other way through the pin-hole
(200 μm).
6. Gently rotate the FOA around the wire to ensure that no particle
remains.
7. Finally, whenever the proper instrumentation is available, a visual
camera with enough magnification might allow to inspect the
cleanliness of the FOA pin-hole.