EXFO IQS-12002B Calibration System User Manual
Page 62
 
Calibration Methods
58
IQS-12002B
Power Meter
At all wavelengths, the output power of the laser sources is measured 
several times alternating between the calibration power meter and the 
power meter under test (DUT). A minimum of five responsivity constants 
are calculated for the DUT and are statistically analyzed to ensure that they 
are within appropriate limits (e.g. the calculated values must all be inside a 
certain range of values). Whenever one or more responsivity constants are 
outside the acceptable range, the measured values are considered invalid 
and the measurements have to be redone.
The number of measurements taken with the calibration power meter and 
the DUT, and the number of comparisons between the two units are set so 
as to reduce uncertainties of Type A. These uncertainties include 
connection uncertainty and interference due to reflections.
The system uses expanded uncertainty (U) values that are included in the 
report. The following tables present the method used to establish the 
uncertainty for calibration made at four wavelengths.