Calibration methods, Power meter – EXFO IQS-12002B Calibration System User Manual

Page 64

Advertising
background image

Calibration Methods

60

IQS-12002B

Power Meter

influence

uncertainty

Unit

probability

distribution

divisor

standard

uncertainty

sensitivity

coefficient

Unit

uncertainty

contribution

i

Xi

u (xi )

ci

ui (y ) (dB)

1

Factors related to the reference standard

rectangular

1.73205

0.000

1

0.000

2

Incertitude d'étalonnage

2.00

%

normal

2

1.000

1

1.000

3

Ref non-linearity

0.25

%

rectangular

1.73205

0.144

1

0.144

4

Ageing

0.00

%

rectangular

1.73205

0.000

1

0.000

5

Ref temperature dependance

1.00

%/°C

rectangular

1.73205

0.577

0.01

°C (variation)

0.006

6

Ref PDR (½ (max-min))

0.64

%

rectangular

1.73205

0.370

0.1

ratio applicability

0.037

7

Ref Dependance on spectral bandwith

0.00

%

rectangular

1.73205

0.000

1

0.000

8

1/2 Ref's Resolution

0.01

%

rectangular

1.73205

0.006

1

0.006

9

Factors related to the DUT

rectangular

1.73205

0.000

1

0.000

10 DUT temperature dependance

1.00

%/°C

rectangular

1.73205

0.577

1

°C (variation)

0.577

11 DUT's Dependance on spectral bandwith

0.00

%

rectangular

1.73205

0.000

1

0.000

12 DUT's PDR (½ (max-min))

0.35

%

rectangular

1.73205

0.202

0.1

ratio applicability

0.020

13 1/2 DUT's Resolution

0.120

%

rectangular

1.73205

0.069

1

0.069

14

rectangular

1.73205

0.000

1

0.000

15

rectangular

1.73205

0.000

1

0.000

16

Combined effects

rectangular

1.73205

0.000

1

0.000

17 Wavelength dependance DUT versus reference std

1.000

%/nm

rectangular

1.73205

0.577

0.5

nm

0.289

18 Source stability

0.12

%

rectangular

1.73205

0.069

1

0.069

19

rectangular

1.73205

0.000

1

0.000

20

rectangular

1.73205

0.000

1

0.000

21

U-shaped

1.41421

0.000

1

0.000

22

rectangular

1.73205

0.000

1

0.000

23

rectangular

1.73205

0.000

1

0.000

24 TYPE A UNCERTAINTY (includes interferance and FOA)

1.50

%

3 measures

1.73205

0.866

1

0.866

y

1.483

k =

2

U =

3.0

Note 1: If there is a polarisation scrambler, the Detector's PDR can be neglected

S=

5.0

Limit 1=

2.6

Limit 3=

7.4

Power meter calibration at 1310 nm and 1550 nm @ 23 °C ± 1 °C

coverage factor

expanded uncertainty