Db) log 10 – EXFO IQS-12002B Calibration System User Manual

Page 68

Advertising
background image

Calibration Methods

64

IQS-12002B

Power Meter

To measure nonlinearity:
1.
Set the attenuators in the two paths so that the power measured on the

meter is the same whether the light is coming from branch a or from
branch b.

2. Open both shutters and measure the total power from both branches

simultaneously: P

total,i

.

3. Close the shutter on branch b and measure the power on branch a:

P

a,i

.

4. Close the shutter on branch a, open the shutter on branch b, and

measure the power on branch b: P

b,i

.

5. Add the individual power measurements.

If the sum is not equal to the total power measurement, the DUT
presents nonlinearity.

:

6. Using the first attenuator, attenuate the total power by the following

factor to the level of the sum of the individual power measurements
calculated in step 5.

7. Repeat steps 2 through 6 to cover the desired range.

(dB)

log

10

i

b,

i

a,

i

total,

10

i

P

P

P

NL

+

×

=

10

2

log

3,01 dB

Advertising