6

ALS Simulator

Laerdal Recommends

Do not insert needles into blood pressure arm .

Heart, Breath and Bowel Sounds and Speech
The ALS Simulator has heart, breath and bowel sounds, both normal and

abnormal with adjustable instructor controlled volume .
Ensure that the ALS Simulator is connected to the Link Box, via manikin cable
located on lower left side of manikin .

See SimPad DFU for complete Heart, Breath and Bowel Sound

Auscultation and Speech functionality and operating information.

Spontaneous Breathing and Airway Closures
The ALS Simulator has spontaneous breathing (visible chest rise and fall) with

variable breathing rate . The breathing is generated via an enclosed air tank in

the right thigh that is pressurized using a standard air pump. (See “

Filling the

Thigh Air Tank” located in the “Getting Started” section of the DFU for

more information) .

The left lung and right lung can be closed independently or together to create a

partial or complete airway obstruction .

See SimPad DFU for complete breathing and airway blockage

functionality and operating information.

After Use:

Cricothyroid Membrane/ Neck Skin Collar:
1 . Remove neck skin collar, by detaching Velcro

®

.

2 . Remove used Cricothyroid Membrane Tape from cricoid opening .
3 . Cut a two (2) inch strip of Cricothyroid Membrane Tape .
4 . Adhere tape to edges of cricoid opening, covering the opening with tape, to

create the actual membrane .

5 . Lay a Neck Skin Collar into molded track around neck area of the manikin .

(Figure 3)

NOTE: If the used collar is still in good repair simply slide the

collar in either direction for a fresh site . If the collar is showing wear, discard
and replace with a new one .

6 . Attach collar using Velcro

®

strips .

Laerdal Recommends

The neck skin collar is designed to provide multiple sites for needle and

surgical techniques . When a fresh site is needed, slide the collar in either

direction . If the neck skin collar is showing wear, discard and replace with a

new collar .

IV Cannulation

After IV training is complete:
1 . Close both clamps on IV tubing and detach bag tubing from arm tubing . Fill a

100cc syringe with water. Flush IV arm with syringe.

Note: This should be done over a basin to catch any fluid.

2 . Using the empty syringe push air through venous system to remove excess

water .

Intramuscular Injections
After intramuscular injection training is complete:
1 . Remove sponge from inside of injection pad skin .

NOTE: Do not store wet foam pads in skin. This will cause

shrinkage of injection pad skin.

2 . Rinse sponge with water, allowing it to dry overnight .
3 . Insert sponge back into injection pad skin .

NOTE: A small amount of powder

applied to sponge will ease insertion .

4 . Replace the pad in the manikin .

Urinary Catheterization (if genitalia upgrade kit purchased)
After urinary catheterization is complete:
1 . Remove complete genitalia assembly .
2 . Working over a basin empty the reservoir .
3 . Rinse thoroughly and allow to dry overnight .

Pneumothorax:

(A) Bilateral Mid-Clavicular Sites:
1 . Detach the chest skin from torso from tabs at the shoulder and back and

remove .

2 . Remove the hard chest plate from the torso .
3 . Disconnect bladder hose from Y connector, located on the underside of the

chest plate . (Photo 11)

4 . Pinch and remove the bladder through the opening between the 2nd and 3rd

intercostal spaces on the top side of the hard chest plate . (Photo12)

5. Trim bladder tubing to match original tubing length to fit this manikin style.
6 . Insert the new pneumothorax bladder into the top side of the chest plate

though the 2nd and 3rd intercostal space opening with the bladder tubing
exiting though the back side of the chest plate . NOTE: The narrow edge of
the bladder should be inserted closest to the sternum . (Photo 13)

7 . Reconnect hose to Y hose connector .
8 . Return chest plate to proper position on torso .
9 . Replace chest skin over torso, being sure to secure skin at shoulders and back .
10. Fill puncture marks on the exterior of the chest skin with the Bladder

Replacement Kit .

Fig . 3

Photo 11

Photo 12

Photo 13

Advertising