DR Power 3-Point Hitch Rapid-Feed User Manual

Page 34

Advertising
background image

30

3-POINT HITCH DR

®

RAPID-FEED™ CHIPPER

1. Remove the four Screws that secure the Front Access

Cover with a 5/16" wrench and remove the Access
Cover (Figure 35).

2. Rotate the Flywheel with a long stick until the Knife is

visible.

3. If the Knife has visible nicks or damage it must be

sharpened or replaced (see “Removing and Replacing
the Chipper Knife” in this chapter).

4. If the Knife does not appear to have any damage,

replace the Access Cover and secure with four Screws
using a 5/16" wrench.

VISUAL INSPECTION OF THE CHIPPER KNIFE AND WEAR PLATE
EVERY 8-10 HOURS

Tools Needed:

• Two 1/2" Wrenches

1. Have someone support the Hopper as you use two

1/2" wrenches to remove the six Bolts, Washers and
Locknuts that support the Hopper to the Chipper

Assembly (Figure 36).

2. Remove the Hopper from the Chipper Assembly.

3. Use a stick to rotate the Flywheel until the Knife is next

to the Wear Plate.

4. Closely inspect the Chipper Knife and Wear Plate for

nicks or dull (rounded) edges (Figure 37).

5. If necessary, repair or replace the Chipper Knife

and/or Wear Plate per the following procedures. If the
Knife and Wear plate do not need maintenance or
adjustment, replace the Hopper.

Front
Access
Cover

Screws

Figure 35

Bolt, Washer (both
sides) and Locknut
(6 places)

Chipper
Assembly

Hopper
Assembly

Figure 36

Chipper
Knife

Figure 37

Wear
Plate

THE HOPPER SHOULD BE SUPPORTED WHEN THE MOUNTING HARDWARE IS REMOVED.
REMOVAL OF THE HOPPER IS DIFFICULT FOR ONE PERSON. TWO PEOPLE ARE NEEDED FOR THE
HOPPER REMOVAL PROCESS.

Advertising